UT Payload – Annual Calibration and Calibration Certificate
1. Purpose of the Calibration Certificate
The calibration certificate issued with the E3 UT Payload serves three main purposes:
A. Verification of proper functionality
The certificate confirms that the E3 UT Payload has been tested and verified to function correctly before leaving Flyability.
This verification applies to each individual UT Gauge + Probe pair. In this context, the certificate serves as a standard conformity certificate, demonstrating that the equipment has been checked and meets the expected functional requirements.
B. Validation for certified inspections
The certificate provides documented proof that the E3 UT Payload has been tested and its proper functionality validated.
Inspectors can use this document as evidence that the equipment has undergone the necessary checks and that its functionality has been verified. In this context, the certificate serves as a calibration certificate and supports the requirements for carrying out certified inspections.
C. Compliance with annual calibration requirements
UT inspection standards require a UT gauge to be calibrated annually.
To address this requirement, the certificate indicates:
The date of calibration.
The next calibration due date.
This effectively provides a guarantee of measurement accuracy for a one-year period. More importantly, having a calibration performed within the preceding year is a requirement for certified inspections. The calibration certificate can therefore be used as supporting documentation to demonstrate compliance with this requirement.
2. When Does the Calibration Validity Period Start?
The calibration validity period starts on the date Flyability ships the E3 UT Payload, rather than the date the customer receives it or puts it into operation.
This means that the one-year validity period is already running while the payload is being transported, stored or waiting to be delivered to the end customer.
For example, if a distributor receives an E3 UT Payload and keeps it in stock for several months before selling it, the remaining calibration validity period will be reduced accordingly.
3. Why Is the E3 UT Payload Manufactured on a Build-to-Order Basis?
To ensure that the E3 UT Payload is manufactured as close as possible to its delivery date, Flyability has adopted a build-to-order manufacturing process instead of its standard made-for-stock approach.
This means that each UT Payload is manufactured upon order rather than being produced in advance and held in Flyability's inventory.
For distributors and resellers, it is important to consider that the one-year calibration validity period begins on the date Flyability ships the payload.
Consequently, if a UT Payload is kept in distributor or reseller stock before being delivered to the end customer, this storage period is included in the one-year calibration validity period.
4. How to Request an Annual Calibration
When the annual calibration of the Elios 3 UT measurement hardware is due, customers can request recalibration directly through Flyability.
To arrange an annual calibration:
Contact Flyability Support: Submit a request to the Flyability Support team (or open a ticket via the Flyability Customer Portal) indicating that your UT equipment is due for its annual calibration.
Provide equipment details: Include the serial numbers of the UT Gauge and the corresponding UT Probes requiring calibration.
Receive RMA & shipping instructions: Flyability Support will provide a Repair Order (RO) along with packing and shipping instructions for sending the gauge and probes. (Note: Do not ship the entire payload arm or drone).
Calibration procedure: The UT Gauge and Probe pairs undergo standard testing, verification, and recalibration in accordance with manufacturer specifications.
Receive your equipment and Calibration Certificate: Following successful verification, the equipment is returned with an official Calibration Certificate documenting the calibration date, standards used, and the next calibration due date.
For questions regarding calibration turnaround times, pricing, or procedures, please contact Flyability Support
Last updated